DUNLEE PHILIPS DA180 Replacement Tube

DUNLEE PHILIPS DA180 Replacement Tube

Availability: In stock

Quick Overview

Replacement Tube for Philips TOMOSCAN CX and TX CT Systems (#885-00, DA/B 180* 823-02)

  • This product is no longer available
  • Tube Type: DU 1505, 1.5 MHU
  • (1.06 MJ) insert
  • Target Diameter: 5 in. (127 mm)
  • Target Angle: 10°
  • Power Characteristics: 56/130 kW
  • Operational Characteristics: 50/60 Hz 150/180 Hz
  • Focal Spot: 0.8 / 1.5
  • Volts: 3.5-11.0 / 6.0-9.5
  • Amperes: 3.0-5.5 / 3.0-5.5
  • Rotor Speed: 9600 RPM
  • Main Winding: 14 Ohms
  • Secondary Winding: 46 Ohms
  • Max. Anode Cooling Rate: 240 kHU/min.(2.8 kW)
  • Ordering No.989605582302

Details

This 1.5 MHU CT tube is specifically designed as a replacement for use with Philips TOMOSCAN CX* and TOMOSCAN TX* CT systems. The DA 180 uses a housing designed and manufactured by Dunlee. The B 180 is a reload of the original equipment manufacturer’s housing using an insert designed and manufactured by Dunlee.